title: "LabAirTec green ductless fume cupboard"
description: "LabAirTec green ductless fume cupboard: a recirculating, filtered unit with Erlab molecular filtration that needs no duct to the outside. Request a quote."
canonical_url: "https://www.renggli.pl/en/offer/fume-hood/non-extraction-fume-hood"
language: "en"
alternates:
  pl: "https://www.renggli.pl/pl/oferta/dygestoria-laboratoryjne/bezwyciagowe.md"
  de: "https://www.renggli.pl/de/angebot/abzuege/umluftabzug.md"
  uk: "https://www.renggli.pl/uk/propozycja/vytiazha/bezkanalna-vytiazhna-shafa.md"
  lt: "https://www.renggli.pl/lt/pasiulymas/garu-traukiai/garu-spintele-be-istraukimo.md"
---

# LabAirTec green ductless fume cupboard

> LabAirTec green ductless fume cupboard: a recirculating, filtered unit with Erlab molecular filtration that needs no duct to the outside. Request a quote.

![Renggli ductless fume hood with a filtration module, control panel and safety cabinets](https://www.renggli.pl/m/dygestorium-bezwyciagowe-renggli-v1__h1366f.webp)

- Category: [Laboratory fume hoods](https://www.renggli.pl/en/offer/fume-hood)

The LabAirTec green ductless fume cupboard is a recirculating, filtered workstation: Erlab molecular filters clean the air from the chamber and return it to the room, so no duct to the outside is needed. It suits repeatable work with a known list of substances where an extract duct cannot be run.

## Key parameters

- Works without a connection to building ventilation
- The most advanced molecular filtration technology
- Retention capacity increased by 50%, extending the filter's lifespan
- Fume hood activation after registering a personalized RFID card

## Advanced features

### Features of the LABAIRTEC GREEN fume hood:

- **One filter for all**: Effective filtering of acid, solvent, and base vapors using the same filter.
- **Remote monitoring**: This function allows for continuous device monitoring, enabling the administrator to check the fume hood's status in real-time from an external computer without the need to shut it down.
- **Broad spectrum**: Ability to work with most commonly used chemicals due to a wide spectrum of retained particles.
- **Filter check**: The system informs about the presence of the filter, distinguishing between new and used filters, ensuring proper usage.
- **Vapor detection**: Detects solvents and acids, providing full protection for users.
- **Temperature alarm**: Monitors temperature fluctuations inside the fume hood and triggers a sound alarm in case of danger.

The LabAirTec green ductless fume cupboard is a recirculating, filtered workstation: Erlab molecular filters clean the air from the chamber and return it to the room, so no duct to the outside is needed. It suits repeatable work with a known list of substances where an extract duct cannot be run.

LabAirTec "green" fume hoods are Renggli's response to the constantly rising energy and installation costs, as they operate without the need to connect to a ventilation system.

Qualified engineers from Renggli and Erlab SAS have jointly designed unique solutions that are friendly to both the user and the environment. The emission of pollutants expelled through the filters is 100 times lower than the permissible threshold values.

LabAirTec "green" fume hoods also feature innovative communication software for monitoring and managing work safety. This system allows management of a network of up to 250 fume hoods.

## How a ductless fume cupboard works {#how-a-ductless-fume-cupboard-works}

A ductless fume cupboard, also called a recirculating or filtered fume cupboard, draws room air in through the sash opening just like a ducted unit. Instead of sending the air outside, it passes it through filters and returns it to the laboratory: adsorption (molecular) filters hold back vapours and HEPA filters capture particles. Because nothing leaves the building, the unit needs no extract duct and no connection to the ventilation system, and it can be moved to another room.

Protection depends on the filter matching the substances in use. HSE guidance [HSG258](https://www.hse.gov.uk/pubns/books/hsg258.htm) recommends HEPA filters for recirculating fume cupboards that control dust, mist or fume, and adsorption filters where you can predict when the filter is likely to fail. Recirculatory filtration fume cabinets have their own European standard, EN 17242, separate from EN 14175, which covers ducted fume cupboards.

## Ductless or ducted: which to choose {#ductless-or-ducted-which-to-choose}

A ductless fume cupboard fits well where:

- no extract duct can be run, for example in an existing building or a rented space,
- the work is repeatable and the substances and quantities are known in advance,
- the workstation may need to move, for example in a teaching laboratory.

A [ducted LabAirTec fume cupboard](https://www.renggli.pl/en/offer/fume-hood/extraction-fume-hoods) is the better choice when substances change from day to day, when larger quantities evaporate or when local rules restrict recirculation, as they can for carcinogenic, mutagenic and reprotoxic (CMR) substances. Both types are compared in our article on the [carbon-filter fume hood](https://www.renggli.pl/en/blog/carbon-filter-fume-hood), and the full range is on the [laboratory fume cupboards](https://www.renggli.pl/en/offer/fume-hood) page.

## Frequently asked questions

**What is a ductless fume cupboard used for?**

For work with chemicals that give off vapours in rooms where no extract duct can be run. The LabAirTec green cleans the air with Erlab molecular filtration and returns it to the room, so it suits repeatable processes with a known list of substances. Parent category: [laboratory fume cupboards](https://www.renggli.pl/en/offer/fume-hood).

**Which standard covers ductless fume cupboards?**

EN 14175 covers fume cupboards connected to an extract system. Recirculatory filtration fume cabinets have a separate standard, EN 17242. Before you buy, ask which filter is fitted and which substances it is suitable for, and check local rules for CMR substances, which can restrict recirculation.