Fume cupboard testing: how often, who tests and what to check
In Great Britain a fume cupboard counts as local exhaust ventilation (LEV), so COSHH regulation 9 requires a thorough examination and test at least every 14 months, with the record kept for 5 years. A competent person carries out the test. Between tests, trained staff run routine checks from the LEV logbook, and users check the test label and the airflow indicator before each use.
Checked:
Prepared by Renggli Sp. z o.o., a laboratory furniture manufacturer whose range includes LabAirTec fume cupboards. The rules and intervals come from the sources listed at the end of the page.
Contents
How often should fume cupboards be tested?
Regulation 9 of the Control of Substances Hazardous to Health Regulations 2002 (COSHH) sets the limit: local exhaust ventilation must have a thorough examination and test at least once every 14 months. HSE guidance HSG258 treats a fume cupboard as a partial enclosure, one of the hood types of an LEV system, so the 14 month limit applies to laboratory fume cupboards in universities, hospitals, industry and schools. Schedule 4 of COSHH sets shorter limits only for a few industrial processes, such as blasting metal castings (1 month) and dry grinding of metal articles (6 months).
Fourteen months is the legal maximum. HSG258 asks for more frequent tests where wear is likely to reduce performance between examinations, for example a cupboard in daily use with corrosive fumes. Booking the test every 12 months leaves two months for a repeat visit after a failed result, or for a lab that cannot be cleared on the planned day.
Who can test a fume cupboard?
HSE expects a competent person, and HSG258 explains what that means: someone with the skills, experience and knowledge to judge the LEV objectively. It can be an outside contractor or a competent employee of the organisation that owns the fume cupboard. Before the visit, clear chemicals and apparatus out of the cupboard and clean the work surface, so the examiner can work on it safely.
The examiner leaves a written record. Keep it, together with any repairs that followed, for at least 5 years. HSG258 also suggests a label on each hood with the test date, the next test date and the examiner's name, and a red fail label when a hood no longer controls exposure: reduced or no detectable airflow, or a cupboard that does not contain the contaminant cloud. Defects should be put right as soon as possible or within the time the examiner sets.
What the thorough examination and test covers
For booths and fume cupboards, HSG258 starts with face velocity: the examiner measures the air speed across the sash opening, and the readings should not vary excessively. It adds that fume cupboards should be further tested to the appropriate British or European standard. In the UK that is BS EN 14175-4, the on-site test methods of the EN 14175 series. Part 4 is written for the place of installation and covers commissioning after installation, routine maintenance tests and the qualification of cupboards that were never type tested. The owner or user decides which methods to apply, and a routine test uses fewer of them than commissioning.
The quantities Part 4 can check include face velocity, containment at the sash opening, robustness against disturbance, pressure drop, sash safety and sound pressure level. Part 4 does not cover recirculating filter units or fume cupboards for radioactive work. For a recirculating cupboard that controls dust, mist or fume, HSG258 asks for a HEPA filter, a check of the filter seating at every filter change and continuous monitoring.
A face velocity figure means little without a benchmark. HSG258 expects a commissioning report when the system is new, with benchmark readings copied into the LEV logbook, so each later test is compared with how the cupboard performed when it was installed.
What to check, how often and who does it
| Check | How often | Who | Record | Source |
|---|---|---|---|---|
| Thorough examination and test of the fume cupboard as LEV | At least every 14 months | Competent person: contractor or trained employee | Written record, kept for at least 5 years | COSHH regulation 9 |
| Commissioning test of a new cupboard (face velocity, flow rates, pressures) | Once, before first use | LEV commissioner or installer | Commissioning report; benchmark readings in the logbook | HSG258, para. 277 |
| Test label on the cupboard (test date, next test, examiner) | At every examination | Examiner | Label on the cupboard; red fail label if control is inadequate | HSG258, paras. 345 and 346 |
| Routine checks: airflow indicator, sash, damage, blockages | Daily, weekly or monthly, as set in the logbook | Trained employee | LEV logbook with the name of the person who checked | HSG258, paras. 318, 319 and 324 |
| Recirculating cupboard controlling dust, mist or fume: filter seating | At every filter change, plus continuous monitoring | Trained employee or service engineer | LEV logbook | HSG258, para. 235 |
| School fume cupboard: test sticker and airflow indicator | Before each use | Teacher or pupil | Sticker showing a pass within 14 months | CLEAPSS SSS100 |
Intervals are legal maximums or guidance minimums. Your risk assessment and the manufacturer's instructions can call for more frequent checks.
Checks between examinations
HSG258 expects the LEV logbook to hold a schedule of daily, weekly and monthly checks, their results and the name of the person who made them. A trained employee can do these checks. For a fume cupboard they are short: the airflow indicator shows normal flow, the sash runs smoothly and stays where it is set, the baffle slots are clear and nothing in the cupboard is damaged. Report any fault to whoever is responsible for the LEV. Cleaning and day-to-day care are covered in our article on fume hood maintenance.
School fume cupboards
Schools follow the same 14 month rule. CLEAPSS guide G9, Fume Cupboards in Schools, covers choosing, installing, maintaining and testing them. G9 replaced Building Bulletin 88, and its 2025 version is a minor revision that updates references without changing the design or performance requirements. CLEAPSS student safety sheet 100 tells pupils to look for a sticker or card showing that the cupboard passed its LEV test within the last 14 months, and to check that the airflow indicator shows safe levels once the fan is on. Furniture for a school lab is described on our school science labs page.
Type test, commissioning test and routine test
The three tests answer different questions. A type test to BS EN 14175-3 checks one model of fume cupboard in a test laboratory, at a stated airflow. A commissioning test to Part 4 checks the installed cupboard in your lab, connected to your extract system. A routine test repeats a reduced set of those measurements at each examination. On-site results cannot be transferred to other cupboards of the same model, which is why manufacturers base their declarations on the type test, and a type test report says nothing about how well a cupboard is connected and balanced in your building. All parts of EN 14175 are explained on our standards page.
Test data for the fume cupboards we supply
For the T120, T150 and T180 bench fume cupboards (Renggli AG), the type test reports and certificates are on our downloads page. The type tests to EN 14175-3 were carried out at the independent Tintschl laboratory in Erlangen; the reports from 2006 give an exhaust airflow of 480, 600 and 720 m³/h at face velocities of 0.34, 0.27 and 0.27 m/s. Give these figures to whoever commissions the cupboard, so the on-site readings are compared with the airflow at which the model was tested. Renggli AG also issued a declaration of conformity with EN 14175 parts 1 to 6 for its Euro II and LabAirTec fume cupboards on 6 February 2015.
We deliver fume cupboards with exhaust ducts, ready for connection; the extract fan is not part of the supply. The face velocity measured on site therefore depends on the building's extract system as much as on the cupboard. Ducted and ductless models are listed in our fume cupboard range, and our service responds to a reported fault within 48 hours.
Frequently asked questions
In Great Britain at least every 14 months: COSHH regulation 9 sets this limit for local exhaust ventilation, and a fume cupboard is LEV. Booking it every 12 months leaves a margin for a repeat visit. Between tests, trained staff run routine checks from the LEV logbook, and users check the test label and airflow indicator before each use.
Yes. HSE guidance HSG258 lists the fume cupboard as a partial enclosure, one of the hood types of an LEV system, and describes face velocity measurement for booths and fume cupboards. That puts laboratory fume cupboards under the COSHH duty to maintain, examine and test LEV.
BS EN 14175-4, the on-site test methods of the EN 14175 series. It is used for commissioning after installation, for routine tests and for qualifying cupboards that were never type tested. The requirements themselves are in BS EN 14175-2, and the laboratory type test is in BS EN 14175-3.
A competent person: someone with the skills, experience and knowledge to judge the LEV objectively. HSG258 says this can be an outside contractor or a competent employee of the organisation that owns the cupboard. The examiner gives you a written record, which you keep for at least 5 years.
HSG258 suggests a red fail label on the hood when airflow is reduced or undetectable, or when the cupboard does not contain the contaminant cloud. Defects should be put right as soon as possible or within the time set by the examiner, and the cupboard should not be relied on for protection until it passes again.
Yes, at least every 14 months like any other LEV. CLEAPSS guide G9, Fume Cupboards in Schools, covers maintaining and testing them, and CLEAPSS student safety sheet 100 tells pupils to check for a sticker showing a pass within the last 14 months before using the cupboard.
Sources
- 1.The Control of Substances Hazardous to Health Regulations 2002, regulation 9, legislation.gov.uk, LEV: thorough examination and test at least every 14 months; records kept for 5 years
- 2.The Control of Substances Hazardous to Health Regulations 2002, Schedule 4, legislation.gov.uk, shorter intervals for four industrial processes
- 3.HSG258 Controlling airborne contaminants at work: A guide to local exhaust ventilation (LEV), third edition, 2017, Health and Safety Executive, paras. 235, 277, 318, 319, 324, 327, 329, 345, 346 and 354
- 4.G9 Fume Cupboards in Schools (2025 version), CLEAPSS, replaces Building Bulletin 88; copies for non-members on request
- 5.Student safety sheet 100: Using a fume cupboard safely, CLEAPSS, test sticker within the last 14 months, airflow indicator
- 6.EN 14175-4:2004 Fume cupboards, Part 4: On-site test methods (scope), CEN, catalogue entry, commissioning, maintenance and qualification at the place of installation
- 7.DIN EN 14175-4:2004-12 Abzüge, Teil 4: Vor-Ort-Prüfverfahren (German edition of EN 14175-4), DIN Media, measured quantities; results not transferable; recirculating units and radioactive work excluded
- 8.LST EN 14175 series, parts 1 to 8 (Lithuanian national adoption), Lithuanian Accreditation and Standardisation Agency (LASA), search for 14175 in the LASA standards shop