Climatic chamber - application in research and requirements for the room
A climatic chamber maintains a set temperature and humidity to test the stability of medicines, cosmetics and food. Test conditions and room requirements.

Published: Updated Reading time: 9 min read
Contents
A climatic chamber is a laboratory device that holds a set air temperature and humidity for days, weeks or months to show how a sample changes under those conditions. It is used mainly for stability testing of medicines, cosmetics and food, and for testing how well products resist moisture. A chamber placed on a countertop needs a table matched to its weight, in other words suitable laboratory furniture.
What are climatic chambers used for?#
Climatic chambers are used to test the stability and durability of products under controlled temperature and humidity. The operating manual of a climatic chamber for stability testing (opens in a new tab) gives its intended use as stability testing of pharmaceuticals, cosmetics and food under conditions kept constant over long periods.
In pharmaceuticals, this testing is described in the ICH Q1A(R2) guideline published by the European Medicines Agency (opens in a new tab). A stability study shows how the quality of a medicine changes over time under the influence of temperature, humidity and light. The shelf life and recommended storage conditions are set on that basis.
The second group is environmental testing of components and equipment. The IEC 60068-2-78 standard (opens in a new tab), published in Poland as PN-EN IEC 60068-2-78, describes how to check whether a product will withstand transport, storage and use in high humidity. The specimen stays in the chamber for a specified time at a constant temperature, at high humidity and without condensation of water vapour.
How does a climatic chamber work?#
A climatic chamber heats or cools the air in a closed workspace, and a separate system adds moisture to the air or removes it. The controller compares sensor readings with the set values and continuously adjusts heating, cooling and humidification.
In the chamber described in the manufacturer's manual (opens in a new tab), a compressor provides cooling and a hot steam generator evaporates water from a tank. The steam enters the chamber above the fan and mixes with the circulating air. Dehumidification works by condensing moisture on Peltier modules (electrically powered cooling elements) at the back of the unit.
Humidification narrows the operating range. According to the same manual (opens in a new tab), the chamber operates from +10 to +60 °C with humidity control. Without humidification and lighting, this chamber (opens in a new tab) cools the interior down to −10 °C.
How does a temperature chamber differ from a humidity chamber and an environmental chamber?#
A temperature chamber controls only temperature, while a humidity chamber controls temperature and relative humidity at the same time. The latter is also called a climatic chamber or a temperature and humidity chamber. “Environmental chamber” is the broadest name: it covers equipment that reproduces ambient conditions, including light. Manufacturers use these names in different ways, so when ordering, compare the specifications rather than the name.
| Chamber type | What it controls | Typical use |
|---|---|---|
| Temperature chamber (thermal or heat chamber) | temperature | storing and testing samples at a constant or varying temperature |
| Climatic chamber (humidity chamber, temperature and humidity chamber) | temperature and relative humidity | stability testing of medicines, cosmetics and food; testing products in damp heat |
| Environmental chamber | temperature, humidity and other factors, e.g. light | testing how well products resist storage and use conditions |
What conditions are set in the chamber for stability testing?#
In stability testing of medicines, the chamber holds the conditions set out in the guidelines, and which ones apply depends on the climate of the market the medicine is intended for. According to the ICH Q1A(R2) guideline (opens in a new tab), the long-term study covers at least 12 months of data at the time of submission. The guideline was developed for the regions of the European Union, Japan and the United States.
The WHO guidelines on stability testing (TRS 1010, Annex 10) (opens in a new tab) add conditions for climatic zones III and IV, that is, hot and dry countries and hot and humid countries. They also require the equipment to keep conditions within the stated limits, and storage conditions to be monitored and recorded. The effects of excursions after an equipment failure are assessed and reported if they may affect the results, so the chamber record is part of the study documentation.
The table summarises the general conditions for drug substances from the ICH Q1A(R2) guideline and the WHO guidelines. Products stored in a refrigerator or freezer have separate conditions in these documents.
| Study | Chamber conditions | Minimum data period at submission |
|---|---|---|
| Long-term | 25 °C ± 2 °C, 60% RH ± 5% RH | 12 months |
| Long-term, alternative | 30 °C ± 2 °C, 65% RH ± 5% RH | 12 months |
| Long-term, alternative from the WHO guidelines | 30 °C ± 2 °C, 75% RH ± 5% RH | 12 or 6 months, according to the WHO guidelines |
| Intermediate | 30 °C ± 2 °C, 65% RH ± 5% RH | 6 months |
| Accelerated | 40 °C ± 2 °C, 75% RH ± 5% RH | 6 months |
What room requirements does a climatic chamber have?#
A chamber needs a room with a stable ambient temperature, a level and load-bearing surface, and clear space around the housing. The exact values are given in the manual for the specific model, so the manual is used when designing the room and the furniture. The table shows the requirements from the operating manual of a climatic chamber for stability testing (opens in a new tab).
| Requirement | Value in the manual |
|---|---|
| Ambient temperature | from 15 to 28 °C, up to 34 °C with a restricted operating range |
| Room air humidity | 70% at most, non-condensing |
| Distance from the back of the chamber to the room wall | at least 15 cm |
| Distance from the ceiling | at least 20 cm |
| Distance at the sides from walls and other equipment | at least 5 cm |
| Surface | flat, level, able to carry the weight of the chamber, heat-resistant and non-flammable |
| Fixing | anti-tip bracket screwed to the wall |
| Water for humidification | demineralised or distilled, conductivity from about 1 to 10 µS/cm, pH from 5 to 7, chlorine-free |
| Power supply | 230 V or 115 V socket, depending on the version |
The net weight of the chambers in this series (opens in a new tab) ranges from 114 to 249 kg, depending on size. The weight of shelves and samples is added to this, and the total is compared with the load capacity of the table or the floor slab. The manual (opens in a new tab) advises against ultrapure water with a conductivity below about 1 µS/cm, because it corrodes the metal parts of the chamber. The chamber is not placed in an explosion hazard zone or near sources of dust and aggressive vapours.
How do you prepare laboratory furniture for a climatic chamber?#
A climatic chamber is placed on a table with a load capacity matched to the weight of the unit with its samples, against a wall it can be screwed to. We do not make chambers, but we design and build the tables, countertops and fitted units that the chamber works with. According to the Renggli FAQ, a standard table carries about 200 kg per running metre, and heavier equipment needs reinforcement. The steel frames in our systems have a load capacity of up to 300 kg per running metre.
The chamber manual requires a heat-resistant and non-flammable surface. For the chamber, we propose a monolithic ceramic countertop, which withstands high temperatures and concentrated acids. Other materials are compared on our laboratory countertops page. When the chamber is to stand in a recess or an unusual spot in a run of furniture, we design the RL Indywidual system from scratch.
The clearances from the chamber manual are part of the furniture layout design from the start. During installation, we level the furniture on site, and once it is installed we issue a handover and acceptance report. How to choose a table for a balance, microscope or analyser is covered in our article on laboratory tables for instruments.
What to settle before ordering a table for the chamber:
- State the chamber model, its external dimensions and its net weight.
- Add the weight of shelves and samples at full load.
- Mark on the floor plan the wall for fixing the chamber and the clearances from its manual.
- List the utilities: power socket, water tank or water connection, drain.
- Check in the manual whether the chamber can stand on a countertop or needs a base.
Frequently asked questions#
Climatic chambers are used for stability testing under controlled temperature and humidity. The oil and water phases are homogenised in a homogeniser, and raw materials are sterilised in sterilisation equipment.
A temperature chamber controls only temperature. A climatic chamber controls temperature and relative humidity at the same time, so it has a humidification and dehumidification system.
The operating manual of a climatic chamber for stability testing (opens in a new tab) recommends calibration once a year. Temperature calibration requires a calibrated reference instrument. In stability testing, the conditions in the chamber are also monitored and recorded throughout the study.
An air-conditioning chamber is a section of an air handling unit that treats the air supplied to a building. A climatic chamber is a test device that holds set conditions around samples in a closed workspace.
A light chamber is a chamber with lighting for testing the effect of light on a product, e.g. the photostability of medicines. The ICH Q1A(R2) guideline (opens in a new tab) refers to the separate ICH Q1B guideline on this.