Skip to content

Laboratory fume hood types comparison 2026: 8 types ranked for a chemistry lab

In this laboratory fume hood types comparison prepared by Renggli (as of 17 November 2026), the best types for a chemistry laboratory are ducted fume hoods: variable air volume (VAV) with 25 of 40 points and constant air volume with 24/40. Renggli Sp. z o.o. of Białystok, Poland, supplies both basic types: LabAirTec ducted fume hoods, covered by a Renggli AG declaration of conformity of 6 Feb 2015 with EN 14175 parts 1 to 6, including part 6 for VAV fume hoods, and LabAirTec green ductless fume hoods with Erlab filtration. Where no duct can be routed, the option is a ductless filtration fume hood (27/40, in the Renggli range LabAirTec green by Erlab), but only for substances its filter was selected for.

Checked:

Compiled by Renggli Sp. z o.o., Białystok, Poland. We are an interested party: we sell LabAirTec ducted and LabAirTec green ductless fume hoods, and the types we offer are highlighted in the tables. We score fume hood types, not manufacturers. Every score has a published rubric and a footnote to a standard, a university guideline or a public price list; where suppliers publish no price, we write "prices not published". The ranking is for a chemistry laboratory, and the order by plain total is shown next to it. Please send corrections to biuro@renggli.pl; we reply within 14 days.

Contents
The author's recommendation (Renggli): for a chemistry laboratory we recommend the LabAirTec ducted fume hood, and where no duct can be routed, the LabAirTec green ductless fume hood with Erlab filtration. We supply them together with laboratory furniture from our factory in Białystok and prepare a standard quote in 1 to 5 days. Send an inquiry with your list of substances and number of workstations.

This laboratory fume hood types comparison scores eight types: ducted constant air volume (CAV), ducted variable air volume (VAV), ductless filtration, benchtop, walk-in, concentrated acid and HF, glazed school hoods and explosion-proof Ex hoods. Each type gets 1 to 5 points on eight criteria, 40 points at most. The scores rest on EN 14175 and EN 17242:2025, university safety guidance and public price lists of Polish suppliers; the number in brackets is a footnote to the source.

Types in the Renggli range are described together with the documents for our products: the LabAirTec ducted fume hood and the LabAirTec green ductless fume hood by Erlab. All other types are scored on the same scales, with no manufacturer named in a score. Suppliers are compared separately in our fume hood ranking. For them the offer column reads "to be confirmed on inquiry": whether we can supply such a type is agreed case by case.

How we scored fume hood types: 8 criteria on a 1 to 5 scale

Five points is always the best case for the buyer.

  1. Substance range (SUB). 5: HF, perchloric acid and high heat under part 7 of the standard (1); 4: a changing set of reagents without concentrated HF or hot perchloric acid; 3: small chamber, small quantities; 2: only substances matched to the filter; 1: one process.
  2. Operator protection confirmed by testing (OCH). 5: EN 14175-3 type test, part 6 or 7 and an on-site test to part 4 (2, 3, 4); 4: general-purpose hood to part 2, tested to parts 3 and 4 (5); 3: test to EN 17242:2025, protection depends on the filter (6); 2: general standard only; 1: no test standard.
  3. Purchase cost (ZAK), public net prices in Poland. 5: up to PLN 12,000; 4: up to PLN 26,000; 3: up to PLN 45,000 or suppliers publish no prices; 2: over PLN 45,000; 1: over PLN 75,000.
  4. Installation cost (INS). 5: no duct, no fan; 4: short duct, low airflow; 3: duct and fan for a small chamber; 2: duct, fan and supply air balance, plus controls for VAV; 1: plus wash-down, Ex rating or very high airflow.
  5. Running cost (EKS). 5: no filters and no heated make-up air; 4: airflow drops with the sash closed; 3: low airflow or filter changes; 2: constant full airflow; 1: constant, very high airflow.
  6. Building requirements (BUD). 5: a socket and general ventilation; 4: short duct, no building works; 3: duct and a bench rated for about 150 kg, or a glazed station visible from the room; 2: duct to the roof, fan and supply air matched to exhaust; 1: plus wash-down drain, Ex zone or room for tall equipment.
  7. Placement flexibility (UST). 5: movable with no works and no new risk assessment; 4: movable after a new risk assessment; 3: fixed place, easy to dismantle; 2: fixed place at the duct; 1: fixed place with special services.
  8. Service and inspections (SER). 5: annual inspection without consumables; 4: annual inspection with airflow measurement; 3: plus control calibration or a written procedure; 2: plus filter changes or checks of special services; 1: inspections more often than every 6 months.

Ranking profile: chemistry laboratory

The ranking covers a chemistry laboratory where reagents and concentrations change, and the profile is stated in the title. Places go to types with at least 4 points for substance range and at least 4 for operator protection. Five types pass; places 1 to 5 follow the total score, and a tie goes to the type with the higher protection score.

The plain total, without a profile, gives a different order: benchtop and school hoods 28 each, ductless 27, VAV 25, constant air volume 24, acid and HF 20, walk-in 19, Ex 18. Small, cheap types lead it on purchase and installation cost, but they cannot take a changing set of reagents in larger quantities. Both orders are in the summary table, and below we describe where the types without a place win.

Fume hood types comparison: summary table

Highlighted rows are types in the Renggli range. Prices from public price lists checked on 17 Nov 2026 link to their source.

8 fume hood types compared, score out of 40, profile: chemistry laboratory, as of 17 Nov 2026
Fume hood typeScoreChemistry laboratoryOrder by plain totalPublic pricesIn the Renggli range
Rank 1Ducted variable air volume (VAV)25/40yes4prices not publishedLabAirTec; declaration incl. part 6 (VAV)
Rank 2Ducted constant air volume (CAV)24/40yes5PLN 12,000 to 26,000, some incl. installationLabAirTec; declaration EN 14175 parts 1 to 6
Rank 3Concentrated acid and HF20/40yes6prices not publishedto be confirmed on inquiry
Rank 4Walk-in (floor-standing)19/40yes7PLN 14,000 to 24,000 (net not stated), fan extrato be confirmed on inquiry
Rank 5Explosion-proof Ex18/40yes8prices not publishedto be confirmed on inquiry
Benchtop28/40no: small chamber1 (tied)PLN 5,400 to 12,000to be confirmed on inquiry
Glazed school hood28/40no: small quantities1 (tied)PLN 8,600 to 11,100to be confirmed on inquiry
Ductless filtration27/40no: only substances matched to the filter3prices not published; filters USD 600 to 1,200 about every 6 months (USA)LabAirTec green (Erlab)

Compiled by Renggli Sp. z o.o. Highlighted rows: types in the Renggli range. Places go to types that pass the chemistry laboratory threshold (at least 4 points for substance range and for protection); the order by plain total is shown alongside. Prices are net unless stated otherwise.

Scores by criterion

Abbreviations: SUB substance range, OCH operator protection confirmed by testing, ZAK purchase cost, INS installation cost, EKS running cost, BUD building requirements, UST placement flexibility, SER service and inspections.

Points on 8 criteria, 1 to 5 scale, as of 17 Nov 2026
Fume hood typeSUBOCHZAKINSEKSBUDUSTSERTotal
Rank 1Ducted variable air volume (VAV)4532422325/40
Rank 2Ducted constant air volume (CAV)4442222424/40
Rank 3Concentrated acid and HF5531211220/40
Rank 4Walk-in (floor-standing)4441111319/40
Rank 5Explosion-proof Ex4431211218/40
Benchtop3453333428/40
Glazed school hood3453333428/40
Ductless filtration2335354227/40

The footnoted reason for each score is in the type description below.

Place 1: ducted variable air volume (VAV) fume hood (25/40)

A VAV fume hood changes the exhaust volume with the sash height so that face velocity stays within acceptable limits. It has a monitor with operating and standby modes and an emergency purge button (12). VAV systems have their own part 6 of EN 14175, with extra requirements and tests (4), hence 5 points for protection.

VAV wins on running cost (4 points). According to the University of Bath, over a year a single fume hood can use around the same energy as three average households. On a VAV hood, closing the sash cuts this by 40 to 75%, and a well-run "shut the sash" practice saves around GBP 1,000 a year per fume hood (13). The University of Edinburgh puts VAV savings at up to GBP 1,000 a year per hood (14). I2SL recommends VAV for all laboratories except those with fewer than one fume hood per about 200 m² (15).

Limitations. VAV needs technically proficient design, installation and maintenance (12), and supply air has to follow the changing exhaust (2 points each for installation and building). Under high heat load the standard names constant airflow as the safer option (1), so VAV is not the first choice for hot digestion. We found no public prices for VAV fume hoods in Poland (3 points).

In the Renggli range: the LabAirTec ducted fume hood. We deliver LabAirTec ducted fume hoods with exhaust ducts, ready for connection; the exhaust fan is not included, so it is planned as part of the building's ventilation. On 6 Feb 2015 Renggli AG issued a declaration of conformity for Euro II and LabAirTec fume hoods with EN 14175 parts 1 to 6 (16); part 6 sets requirements and tests for variable air volume fume hoods (4). Whether a given LabAirTec configuration runs on variable airflow is confirmed in our quotation. Versions and equipment are described on the LabAirTec ducted fume hood page, and you can get a price for a VAV configuration through an inquiry.

Place 2: ducted constant air volume (CAV) fume hood (24/40)

A CAV fume hood exhausts a constant volume of air, so face velocity rises as the sash comes down. It can have a bypass, an extra air inlet (17). It is the simplest ducted type, and several suppliers publish prices: PLN 12,000 net for a 1200 mm hood and PLN 15,000 net for 1500 mm, both including exhaust installation (Laweko-Plast, Bydgoszcz) (7), and PLN 14,000 to 24,000 net for an aluminium version and PLN 17,000 to 26,000 net for a ceramic worktop version, without installation (POLFLEX, Toruń) (18, 19). Hence 4 points for purchase cost.

Protection is confirmed by an EN 14175-3 type test and an on-site test to part 4 (2, 3). In Poland these tests are carried out by, among others, CIOP-PIB; according to the institute's 2018 information sheet, testing one fume hood takes 3 to 4 weeks (20).

Limitations. Constant full airflow costs money: the hood exhausts heated air the whole time it runs, even with the sash closed (2 points for running cost) (17, 21). For scale, one fume hood running around the clock uses as much energy as a house (Lawrence Berkeley National Laboratory, USA, 2001) (21). The University of Bath advises replacing CAV hoods with VAV models when buying, where the building services allow (13). Exhaust flow is sash opening area times face velocity: a 1.2 m sash opened 0.5 m at 0.5 m/s gives 1,080 m³/h, and the recommended face velocity is 0.3 to 0.5 m/s (22, 17). One fume hood is usually designed for 400 to 1,200 m³/h, with supply air set about 10% below exhaust (22). The calculation for your own sash is in our post on fume hood airflow.

In the Renggli range: the LabAirTec ducted fume hood. LabAirTec ducted fume hoods, with the same declaration of conformity with EN 14175 parts 1 to 6 (16). We publish type test reports separately, for three Renggli AG bench models of the EURO-II line: the T120, T150 and T180 passed an EN 14175-3 type test at the Tintschl laboratory in Erlangen, at constant airflow (23, 24, 25). The certificates and full reports are on our downloads page, and we match the model to your lab in a request for quotation. Fume hoods come with a 24-month warranty that can be extended, and our service responds within 48 hours.

Renggli AG model

Width

Airflow in test

Face velocity

T120

1200 mm

480 m³/h

0.34 m/s

T150

1500 mm

600 m³/h

0.27 m/s

T180

1800 mm

720 m³/h

0.27 m/s

Place 3: concentrated acid and HF fume hood (20/40)

This type has the top scores for substance range and protection, 5 points each. EN 14175-7 covers fume hoods for high heat load, hot acid digestion, perchloric acid and hydrofluoric acid (1). A conventional fume hood should not be used for large quantities of concentrated HF: it needs a polypropylene worktop, a wash-down system and a PVC or other non-glass sash. Heating perchloric acid needs a fume hood with wash-down, because perchlorates build up in the ducts and can explode (12, 17).

Limitations. Wash-down needs a drain, acid-resistant materials and a fixed place (1 point each for installation, building and placement). Under high heat load the standard recommends constant airflow (1), which raises energy cost (2 points). We found no public prices in Poland, so like every type without prices it gets 3 points for purchase. Chamber and seal materials are covered in our post on the perchloric acid fume hood.

In the Renggli range: the LabAirTec ducted fume hood in a perchloric acid version, on inquiry. The LabAirTec ducted fume hood page lists, among the price factors, a version for special substances such as perchloric acid, with a chemically resistant chamber and water wash-down (22). Whether a configuration also covers concentrated HF is confirmed in our quotation, so add the acids, their concentrations and temperatures to your inquiry.

Place 4: walk-in fume hood (19/40)

A walk-in fume hood stands on the floor, has no worktop and takes large apparatus; working in it requires a written operating protocol (17). A market example: a 2100 mm high chamber, a sash that comes down to 50 mm above the floor, widths from 600 to 1800 mm and a price of PLN 14,000 to 24,000, for which the page does not say whether it is net or gross. A ventilation firm selects the fan separately (POLFLEX, Toruń) (8). Hence 4 points for purchase. Protection is confirmed by tests to parts 3 and 4 of the standard (4 points) (2, 3).

Limitations. Exhaust flow is sash opening area times face velocity (22), so a large opening at the same velocity means a large airflow: a large fan and large supply air. This gives the highest energy cost in this comparison; that is our conclusion from the formula, with no separate numeric source. Walk-in hoods therefore get 1 point each for installation, running cost, building and placement.

Place 5: explosion-proof Ex fume hood (18/40)

The Ex version extends use to explosion hazard zones. Equipment in such zones falls under the ATEX Directive 2014/34/EU, in force since 20 Apr 2016 and implemented in Poland by a regulation in Dz.U. 2016 item 817 (26). In practice this means Ex-rated lighting, switches, sensors, sockets and fan, which manufacturers list as options (27). The chamber and airflow testing are the same as for a general-purpose hood (4 points for protection) (2).

Limitations. Zone classification, Ex installation and a fixed place give 1 point each for installation, building and placement. We found no public prices (3 points for purchase). Running cost and service (2 points each) are scored as for constant airflow plus checks of the Ex rating; we have no neutral numeric source for these two scores.

Outside the profile: benchtop fume hood (28/40)

A benchtop fume hood sits on an existing bench, and laboratories short of space choose it. A market example: widths from 600 to 1800 mm, about 150 kg, so the bench needs a matching load rating, a 150 or 200 mm duct and a fan selected separately. Prices in the price lists we checked: from PLN 5,400 net (Pro-Cezas, Zielona Góra) (9) to PLN 12,000 net (POLFLEX, Toruń, PLN 8,000 to 12,000) (27), hence 5 points for purchase.

Why no place. The small chamber takes small quantities of reagents, so substance range gets 3 points and the type does not pass the chemistry laboratory threshold. On the plain total it shares first place with the school hood. It suits a small analytical lab.

Outside the profile: glazed school fume hood (28/40)

In a school lab the fume hood protects the teacher and pupils during demonstrations. Our fume hood category page lists what counts here: glazing on several sides, a sash lock, a simple panel and a position that does not block the board (28); we found no neutral Polish source with such requirements. Hoods in the school category of the Pro-Cezas online shop (Zielona Góra) cost PLN 8,600 to 11,100 net, delivery extra (10), hence 5 points for purchase. The Polish school health and safety regulation (consolidated text Dz.U. 2020 item 1166) does not mention fume hoods; it requires hazardous substances to be stored in suitable rooms and safety data sheets to be available (29).

Why no place. Demonstrations use small quantities (3 points for range). Running cost and placement scores (3 points each) are our assumption for occasional use, without a neutral numeric source. What to look for when buying for a school is covered in our post on the school fume hood.

Outside the profile: ductless filtration fume hood, where there is no duct (27/40)

A ductless fume hood cleans the air with a filter, usually carbon, and returns it to the room, so it needs no duct and no fan (12, 17). It is the only type with 5 points for both installation and building. Moving it only requires a new risk assessment (4 points for placement) (12). Design requirements and type and on-site tests for these units are set out in EN 17242:2025, approved on 6 Nov 2025 (6).

Why no place in the chemistry profile. Protection depends entirely on the filter. Once it is saturated, the substance enters the room, potentially without warning, so the filter must match the substance at a known concentration, and sensors warning of pending breakthrough are recommended (12). Oregon State University advises against this type for multiple chemicals, changing processes and heated reactions, and puts filter replacement at USD 600 to 1,200 roughly every 6 months, with spent filters disposed of as hazardous waste (11). UC Irvine allows it for a consistent process with small quantities of known chemicals (30). Hence 2 points each for substance range and service.

In the Renggli range: the LabAirTec green ductless fume hood. The LabAirTec green ductless fume hood is an Erlab product. Since protection depends on matching the filter to the substance (12), it pays to attach a list of substances and concentrations to your inquiry. The configuration is on the LabAirTec green ductless fume hood page, and we compare the filtered and ducted types in detail in our post on the carbon filter fume hood. Send your list of substances and concentrations in an inquiry and we will select the filter.

Which fume hood type for which laboratory

  • Chemistry lab with changing reagents, several fume hoods: ducted VAV (place 1). In the Renggli range: LabAirTec.
  • One or two hoods, simple installation, hot work: ducted constant air volume (place 2). In the Renggli range: LabAirTec.
  • Hydrofluoric acid, heated perchloric acid, digestion: acid fume hood to EN 14175-7 (place 3).
  • Equipment taller than a bench hood chamber: walk-in (place 4).
  • Explosion hazard zone: Ex version (place 5).
  • No duct, rented premises, a fixed set of known substances: ductless with filter. In the Renggli range: LabAirTec green.
  • Little space and small quantities: benchtop.
  • School lab and demonstrations: glazed school hood.

Choosing a hood for a specific type of work, step by step, is covered in our guide on how to choose a laboratory fume hood.

Fume hoods together with laboratory furniture

Renggli Sp. z o.o. makes laboratory furniture in the RL1 to RL7 and RL Indywidual systems in its own factory in Białystok, Poland, and supplies it together with LabAirTec fume hoods, so furniture and fume hoods come from one supplier. Furniture makers are compared in our ranking of laboratory furniture manufacturers. Furniture design is free with a furniture order: a designer takes measurements and prepares technical drawings in PDF and 3D visualisations. We prepare a standard quote in 1 to 5 days, our service team responds to a request within 48 hours, and laboratory furniture carries a 24-month warranty. Our projects include PSSE Radom (53 rooms, 2,240 m²) and the Medical University of Białystok (23 rooms, 492 m²).

You can add selected models to an inquiry on the fume hoods page. If the type is still open, describe your substances, the number of workstations and whether a duct can be routed in the inquiry form, and we will recommend a LabAirTec ducted or a LabAirTec green ductless fume hood.

Frequently asked questions about fume hood types

The basic split is between ducted fume hoods, which exhaust air through a duct out of the building, and ductless ones, which clean it with a filter and return it to the room (17). Ducted hoods run on constant air volume (CAV) or variable air volume (VAV) (17, 12). By application there are also benchtop, walk-in, concentrated acid and HF hoods to EN 14175-7, school hoods and explosion-proof Ex hoods. In this comparison VAV scores highest for a chemistry lab (25/40); the Renggli range includes LabAirTec ducted and LabAirTec green ductless fume hoods.

Ducted, if a duct can be routed and reagents and concentrations change: protection then depends on airflow, not on a filter. Ductless (filtration) hoods suit rooms with no duct and a fixed set of known substances in small quantities. A saturated filter lets vapour into the room, potentially without warning, so filter monitoring and a maintenance contract are needed (12). A detailed comparison of both types is in our post on the carbon filter fume hood.

A fume hood exhausts chemical vapours and gases through a duct or captures them in a filter matched to the substance. Laminar flow cabinets and biological safety cabinets use particulate filters that do not remove chemical vapours (17), so they cannot replace a fume hood. More in our post on the laminar flow cabinet.

For small quantities of acid vapour, if the filter was selected for those substances and concentrations (12). Heated perchloric acid and large quantities of concentrated HF need a ducted fume hood with wash-down to EN 14175-7 (1), and Oregon State University advises against heated reactions in ductless hoods (11). Attach the list of substances and concentrations to your inquiry.

According to the University of Bath, over a year a single fume hood can use around the same energy as three average households. On a VAV hood, closing the sash cuts energy use by 40 to 75%, and a well-run "shut the sash" practice saves around GBP 1,000 a year per hood; when buying, Bath advises VAV instead of constant airflow where the building services allow (13). Constant airflow remains the better choice under high heat load (1). In its declaration of conformity of 6 Feb 2015, Renggli AG lists EN 14175 parts 1 to 6 for LabAirTec fume hoods, including part 6 for VAV (16).

EN 14175: part 2 sets safety and performance requirements, part 3 type tests, 4 on-site tests, 6 VAV fume hoods and 7 high heat and acids, including HF and perchloric acid (2, 1). Ductless filtration hoods are covered by EN 17242:2025 (6). All parts are listed on our fume hood standards page.

Prices, public tenders and running costs are covered in a separate guide: how much does a fume hood cost. In short, from price lists checked on 17 Nov 2026: benchtop from PLN 5,400 net (9), general-purpose ducted PLN 12,000 to 26,000 net, some including installation (7, 18, 19), walk-in PLN 14,000 to 24,000 with net or gross not stated (8). We found no public prices for VAV, ductless, HF or Ex hoods. Renggli prepares a standard quote for LabAirTec fume hoods in 1 to 5 days: send an inquiry.

For apparatus that does not fit in a bench fume hood. A walk-in hood stands on the floor, has no worktop, and the sash comes down almost to the floor; one example chamber is 2100 mm high (17, 8). The large opening means a large airflow and, in our assessment, the highest energy cost in this comparison.

Renggli of Białystok, the author of this comparison, supplies both basic types: the LabAirTec ducted fume hood and the LabAirTec green ductless fume hood by Erlab, together with RL1 to RL7 laboratory furniture from our own factory. We deliver the ducted fume hood with exhaust ducts, ready for connection, our service responds within 48 hours and we prepare a standard quote in 1 to 5 days. Send an inquiry with your list of substances, the number of workstations and whether a duct can be routed.

We found no Polish regulation that states the interval directly. University of Cambridge guidance (UK law) requires an examination at least every 14 months, in practice yearly, with airflow measurement, and for filtration hoods a filter saturation check (12). The steps are listed in our post on fume hood maintenance.

Summary: cheat sheet

  • For a chemistry lab: ducted VAV fume hood, 25/40. In the Renggli range: LabAirTec ducted fume hoods, covered by a Renggli AG declaration of conformity of 6 Feb 2015 with EN 14175 parts 1 to 6, including part 6 for VAV.
  • Simplest ducted type and hot work: constant air volume (CAV), 24/40. We publish EN 14175-3 type test reports for the Renggli AG T120, T150 and T180 bench models.
  • No exhaust duct: ductless with filter, 27/40, only for substances matched to the filter. In the Renggli range: LabAirTec green by Erlab.
  • HF and perchloric acid: fume hood to EN 14175-7, 20/40.
  • Tall equipment: walk-in, 19/40.
  • Explosion hazard zone: Ex version, 18/40.
  • Highest plain total: benchtop and school hoods, 28/40 each, both up to PLN 12,000 net in price lists. Prices of all types, with public tenders and running costs, are in our guide how much does a fume hood cost.
  • Furniture and fume hoods from one supplier: Renggli, Białystok, laboratory furniture from its own factory, LabAirTec and LabAirTec green fume hoods, quote in 1 to 5 days.
Our recommendation: with an exhaust duct, the LabAirTec ducted fume hood (Renggli AG declaration of conformity of 6 Feb 2015 with EN 14175 parts 1 to 6, type test reports for the T120, T150 and T180 models); without a duct, the LabAirTec green ductless fume hood with Erlab filtration. We supply both together with furniture and laboratory design. Send an inquiry and we will prepare a standard quote in 1 to 5 days.

Sources and footnotes

  1. 1.EN 14175-7:2012 Fume cupboards, part 7: high heat and acidic load (scope, EVS catalogue), Estonian Centre for Standardisation and Accreditation (EVS), HF, perchloric acid, high heat; fixed air volume under high heat load
  2. 2.EN 14175-3 Fume cupboards, part 3: type test methods (2019 edition, scope, EVS catalogue), Estonian Centre for Standardisation and Accreditation (EVS), type test; references to parts 6, 7 and EN 17242
  3. 3.EN 14175-4:2004 Fume cupboards, part 4: on-site test methods (scope, EVS catalogue), Estonian Centre for Standardisation and Accreditation (EVS), on-site tests
  4. 4.EN 14175-6:2006 Fume cupboards, part 6: variable air volume fume cupboards (scope, EVS catalogue), Estonian Centre for Standardisation and Accreditation (EVS), requirements and tests for VAV systems
  5. 5.EN 14175-2:2003 Fume cupboards, part 2: safety and performance requirements (scope, EVS catalogue), Estonian Centre for Standardisation and Accreditation (EVS), scope of part 2
  6. 6.SS-EN 17242:2025 Recirculatory filtration fume cabinets (standard record, approved 6 Nov 2025), Swedish Institute for Standards (SIS), standard for ductless filtration fume hoods
  7. 7.Cena dygestorium (price list, in Polish), Laweko-Plast Krzysztof Piasek (Bydgoszcz, Poland), PLN 12,000 net (1200 mm) and PLN 15,000 net (1500 mm) including exhaust installation
  8. 8.Dygestorium walk-in WIN (walk-in fume hood, in Polish), POLFLEX (Toruń, Poland), PLN 14,000 to 24,000 (net or gross not stated), 2100 mm chamber, fan sold separately
  9. 9.Dygestoria nastołowe (benchtop fume hoods, online shop, in Polish), Pro-Cezas (Zielona Góra, Poland), from PLN 5,400 net
  10. 10.Dygestoria szkolne (school fume hoods, online shop, in Polish), Pro-Cezas (Zielona Góra, Poland), Pro-I, Pro-II, Pro-III PLN 8,600 net, Pro-III glass PLN 9,600 net, Pro-IV PLN 11,100 net, delivery extra
  11. 11.Ductless Fume Hood Safety (03/2026), Oregon State University, Environmental Health and Safety, ductless fume hood limits, filter cost (USA)
  12. 12.HSD029C rev 5: Guidance for Safe Use of Laboratory Fume Cupboards (February 2020), University of Cambridge, Health and Safety Division, VAV, recirculating filters, HF, perchloric acid, 14-month examination
  13. 13.Using lab ventilation sustainably, University of Bath, VAV: 40 to 75% less energy with the sash closed, about GBP 1,000 a year
  14. 14.Ventilation best practice (2024), University of Edinburgh, VAV up to GBP 1,000 a year per fume hood
  15. 15.Best Practices Guide: Energy Savings with Demand Based Control, International Institute for Sustainable Laboratories (I2SL), VAV except labs with one hood per more than 200 m²
  16. 16.Renggli AG declaration of conformity: Euro II and LabAirTec fume hoods, EN 14175 parts 1 to 6 (6 Feb 2015), Renggli AG, manufacturer's declaration, in German
  17. 17.A Guide to Laboratory Fume Hoods (June 2015), Tennessee Tech University, Environmental Health and Safety, CAV, bypass, walk-in, perchloric acid, laminar flow cabinets are not fume hoods
  18. 18.Dygestorium aluminiowe ALU (aluminium fume hood, in Polish), POLFLEX (Toruń, Poland), PLN 14,000 to 24,000 net
  19. 19.Dygestorium ceramiczne EKO (ceramic worktop fume hood, in Polish), POLFLEX (Toruń, Poland), PLN 17,000 to 26,000 net
  20. 20.Testing of local exhaust ventilation (PN-EN 14175, PN-EN 1093), information sheet, version 3 of 28 Aug 2018, Central Institute for Labour Protection, National Research Institute (CIOP-PIB, Warsaw), type and on-site testing, 3 to 4 weeks per fume hood
  21. 21.Press release on fume hood energy use (31 Aug 2001), Lawrence Berkeley National Laboratory (USA), one fume hood running 24 hours uses as much energy as a house
  22. 22.LabAirTec ducted laboratory fume hood, Renggli Sp. z o.o. (Białystok), exhaust flow formula and typical airflows (guidance section of the page, in Polish); price factors, including a perchloric acid version with wash-down
  23. 23.EN 14175-3 type test report, Renggli AG EURO-II T120 (1200 mm), Tintschl (Erlangen) for Renggli AG, 480 m³/h constant flow, 0.34 m/s
  24. 24.EN 14175-3 type test report, Renggli AG EURO-II T150 (1500 mm), Tintschl (Erlangen) for Renggli AG, 600 m³/h, 0.27 m/s
  25. 25.EN 14175-3 type test report, Renggli AG EURO-II T180 (1800 mm), Tintschl (Erlangen) for Renggli AG, 720 m³/h, 0.27 m/s
  26. 26.ATEX Directive 2014/34/EU on equipment and protective systems for potentially explosive atmospheres, Polish Ministry of Development and Technology (gov.pl), in force since 20 Apr 2016, Polish implementation Dz.U. 2016 item 817
  27. 27.Dygestorium nablatowe NAL (benchtop fume hood, in Polish), POLFLEX (Toruń, Poland), PLN 8,000 to 12,000 net, about 150 kg, Ø 150 or 200 mm duct, Ex options
  28. 28.Laboratory fume hoods (offer category), Renggli Sp. z o.o. (Białystok), fume hoods in school laboratories (Renggli editorial text, in Polish)
  29. 29.Regulation on health and safety in public and non-public schools, consolidated text Dz.U. 2020 item 1166, Polish Journal of Laws (Sejm), no fume hood requirement; chemical storage and safety data sheets (§ 29)
  30. 30.Position Statement: Ductless Fume Hood Systems (22 March 2011), University of California, Irvine, Environmental Health and Safety, when a ductless fume hood is acceptable
Newsletter

Stay up to date with Renggli

New projects, products and guides on laboratory furniture, in one place.

By subscribing you agree to receive messages from Renggli and accept the privacy policy. You can opt out at any time.

Let us plan your laboratory

Tell us what has to happen in the room. We will pick the furniture and work out the rest.

Quote within 1-5 days, furniture design free with your order.